Features : 1. 30V/5A full digital power control single 2. 10mV/1mA high accuracy and resolution 3. Low noise and ripple 4. CV/CC given voltage and constant current mode 5. Highly reliable OCP, OVP, arbitrarily set protection 6. Built-in set of parameters stored 7. Outputs a switch control 8. Power and memory function 9. Keypad lock to prevent misuse 10. Intelligent temperature control fan, effectively reduce the noise 11. Easy alternative to traditional digital power Precisely generate the required voltages and currents. 0-30V adjustable DC output high-precision. 10mV and 1mA resolution. Professional power supply design with intuitive operability. Simple and intuitive operation, fast and accurate selection of voltage and current can be realized. To prevent damage to the load Can be set to the value of over-voltage protection and over-current protection to prevent the voltage regulator from beyond the threshold can The keyguard lock panel buttons to prevent undesirable results in the key test engineering 5 user-programmable preset settings Just press the button to recall and save area a key to bring up the user to store the data and save user settings Specification : Operating Conditions : Input voltage : 220V /110V+/-10% Frequency range : 50- 60Hz Operating temperature : 32-104 ℉ (0-40 ºC) Operating humidity : < 80%RH Storage temperature : 32-158 ℉ (0-70 ºC) Storage humidity : < 70%RH Stable Voltage : Output Voltage Range : 0-30V Continuously adjustable Setup Resolution : 10 mV Setup Accuracy : <0.05% +20 mV Ripple : < 1mVRMS Read back Accuracy : 10 mV Temp. Coefficient : < 0.1% + 10mV Read back Temp. Coefficient : < 100 ppm+10 mV Stable Current : Output Current Range : 0-5A Continuously adjustable Setup Resolution : 1 mA Setup Accuracy : <0.1% +3 mA Ripple : < 3mARMS Read back Accuracy : 1 mA Temp. Coefficient : < 0.1% + 5mA Read back Temp. Coefficient : < 100 ppm+5 mA Dimensions : 4.3 x 6.1 x 10.25 inch (110 x 156 x 260mm)
Package includes : 1 x KA3005D 0~30V 0~5A Adjustable DC Power Supply 1 x Test line 2 x Probe 1 x Power line(EU/AU/US)
Safety Guidelines
Do not block or obstruct the cooling fan vent opening.
Avoid severe impacts or rough handling that leads to damage.
Do not discharge static electricity.
Do not disassemble unless you are qualified as service personnel.
AC Inut Voltage:
110V/ 120V / 220V / 230V , 50/60 Hz
Connect the protective grounding conductor of the AC power cord to an earth ground, to avoid electrical shock.
Operation Environment
Location: Indoor, no direct sunlight, dust free, almost non-conductive pollution (note below)
Relative Humidity: < 80%
Altitude: < 2000m
Temperature: 0-40°C
Storage environment
Relative Humidity: < 70% •
Temperature:-10-70°C
Location: Indoor
Main Features Performance • Low noise: cooling fan controlled by heatsink temperature
• Compact size, light weight
Constant voltage / constant
Corrent operation
Output On / Off control
• Digital panel control
4 pairs of panel setup save / recall •
Conarse and fine voltage / current control
• Software calibration
• Beep output •
Button lock function
protection
Overload protection
Reverse polarity protection
Interfaces
• USB/RS232 for remote control (only for KA3003P, Operation Protection Interfaces KA3005P, KA6002P, KA6003P, KA6005P & KA3010P)
Condition Indication
OVP
OVPis the indicator of overvoltage protection. When overvoltage function is turned on,
OVP indicator lights on; when output voltage is higher than protection setup value due to unexpected conditions, output cuts off and OVP indicator flickers; Press the button OVP again, and the power supply
OCP
OCP is OCP indicator. When overcurrent function is turned on,
OCP indicator lights on.
C.C
C.C is constant current indicator. When power supply is in the mode of constant current, this light is on.
C.V
C.V is constant voltage indicator. When power supply is in the mode of constant voltage, this light is on.
OUT
OUT OUT is output indicator. If light on, there is voltage output in the output terminal.
Storage Indication
M1
M2 Indication of saving and recalling 4 setups stored internally
M3 When LOCK indication turns on, the front panel button •
M4 operation is locked. •
M5
Brief Introduction of Panel Operation (MEMORY RECALLISAVE ) M1 M2 M3 M4 Saves or recalls panel settings. For settings, 1~4 are available. For save / recall details UNLOCK LOCK Front panel LOCK function. For details, BEEP OCP Over-Current protect on/off,. Pressing this button for more than 2 secends will make beep On/OFF. OVP Over-votage Protect On/Off. Off/ON Output On/Off.